Remote Control Commands
R&S
®
NGP800
113
User Manual 5601.5610.02 ─ 04
[SOURce:]ALIMit[:STATe]
<arg0>[, <Channel list>]
[SOURce:]ALIMit[:STATe]?
[<Channel list>]
Sets or queries the safety limit state.
Parameters:
<state>
1
Activates the safety limit.
0
Deactivates the safety limit.
Parameters for setting and query:
<Channel list>
<list>
Example:
ALIM 1, (@1)
Activates the safety limit state at channel 1
Example:
See
Example "Configuring the safety limit"
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer
<New value for voltage>[,
<Channel list>]
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer?
[<Channel list>]
Sets or queries the lower safety limit for voltage.
Setting parameters:
<voltage>
<numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Numeric value for safety limit.
MIN | MINimum
Min value for lower safety limit.
MAX | MAXimum
Max value for lower safety limit.
Range:
0.000E+00 to 6.4050E+01
Increment: 0.001
*RST:
0.000E+00
Parameters for setting and query:
<Channel list>
<list>
Example:
VOLT:ALIM:LOW? (@1)
Queries the lower safety limit for voltage at channel 1
Example:
See
Example "Configuring the safety limit"
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]
<New value for voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]?
[<Channel list>]
Sets or queries the upper safety limit for voltage.
Setting parameters:
<voltage>
<numeric value> | MIN | MINimum | MAX | MAXimum
Configuration Commands
www.allice.de
Allice Messtechnik GmbH