Remote Control Commands
R&S
®
NGP800
136
User Manual 5601.5610.02 ─ 04
0
OPP is deactivated
1
OPP is activated
Parameters for setting and query:
<Channel list>
<list>
Example:
POW:PROT? (@1)
Queries OPP state at channel 1.
Example:
See
Example "Configuring the overpower protection"
[SOURce:]POWer:PROTection:CLEar
[<Channel list>]
[SOURce:]POWer:PROTection:CLEar
[<Channel list>]
Resets the OPP state of the selected channel. If an OPP event has occurred before,
the reset also erases the message on the display.
Parameters:
<Channel list>
<list>
Example:
POW:PROT:CLE (@1)
Resets OPP state at channel 1.
Example:
See
Example "Configuring the overpower protection"
Usage:
Setting only
[SOURce:]POWer:PROTection:LEVel
[<New value for voltage protection>, <Channel
list>]
[SOURce:]POWer:PROTection:LEVel?
[<Channel list>]
Sets or queries the overvoltage protection value of the selected channel.
Parameters:
<power>
<numeric value> | MIN | MINimum | MAX | MAXimum | DEF |
DEFault
<numeric value>
Numeric value of the power protection level in watts.
MIN | MINimum
Minimum value of the power protection level at 0.00 W.
MAX | MAXimum
Maximum value of the power protection level at 200.00 W.
DEF | DEFault
Default value of the power protection level at 200.00 W.
Range:
0.00 to 200.00
*RST:
200.00
Default unit: W