CALibration:AINPut:UMIN
Sets the output voltage to low value 1 % of Vmax for analog input pin during adjust-
ment.
Usage:
Event
CALibration:CANCel
Cancels the channel adjustment.
Usage:
Event
CALibration:COUNt?
Queries the number of counts channel adjustment performed successfully.
Usage:
Query only
CALibration:CURRent:DATA
<arg0>
Set the DMM reading after setting the output current level in channel adjustment proc-
ess.
Parameters:
<current>
Measured value from DMM.
CALibration:CURRent:IMAX
Sets the output current to high value 100 % of Imax during current adjustment.
Usage:
Event
CALibration:CURRent:IMIN
Sets the output current to low value 1 % of Imax during current adjustment.
Usage:
Event
CALibration:DATE?
Returns the channel adjustment date.
Usage:
Query only
CALibration:END
Ends the channel adjustment.
Usage:
Event
Advanced Operating Commands
www.allice.de
Allice Messtechnik GmbH