Instrument Functions
R&S
®
NGP800
51
User Manual 5601.5610.02 ─ 04
Figure 6-5: Delay text at channel display area
6.2.1.2
Remote Sensing
For voltage setting < 1 V, setting of remote sensing mode to "EXT" is recommended.
The "Remote sensing" is a mechanism used to monitor and compensate the voltage
drops on the cables connected to the load.
1. Select "Remote sensing" menu item to configure the remote sensing mode.
The R&S
NGP800 displays the "Remote sensing" dialog.
2. Select the required remote sensing mode.
● EXT: The internal voltage sense relay in the instrument is switched on and the
connection of remote sensing wires (S+, S- ) to the input of the load become
necessary. Failure to connect remote sense can cause overvoltage or unregu-
lated voltage output from the R&S
NGP800. The voltage sensing relay remains
switched on even when output is turned off.
Activating the Channels Output