Instrument Functions
R&S
®
NGP800
88
User Manual 5601.5610.02 ─ 04
6.16.6
Update Device
Latest instrument firmware is available in the R&S
NGP800 product homepage.
1. Select the "Update Device" to update instrument firmware.
The R&S
NGP800 displays the update device dialog.
Figure 6-39: Update device dialog
2. Select the source and file location to update instrument firmware.
3. Select "UPDATE" to update the instrument firmware.
The R&S
NGP800 updates the instrument firmware accordingly.
6.17
Adjustment
Adjustment shall be done at ambient temperature of 25
°
C
±
2
°
C .
The instrument must be operated for at least 30 minutes before executing the adjust-
ment .
Thick wires are recommended for connecting the shunt resistor to avoid huge voltage
drop and excessive heating.
For ease of maintenance, the R&S
NGP800 provides two adjustment procedures
which you can apply on the instrument:
●
Chapter 6.17.1, "Analog In Adjustment"
Adjustment