Remote Control Commands
R&S
®
NGP800
159
User Manual 5601.5610.02 ─ 04
Parameters:
<voltage>
Measured value from DMM.
CALibration:VOLTage:UMAX
Sets the output voltage to high value 100 % of Vmax during voltage adjustment.
Usage:
Event
CALibration:VOLTage:UMIN
Sets the output voltage to low value 1 % of Vmax during voltage adjustment.
Usage:
Event
7.8
Data and File Management Commands
The
DATA
and
HCOPy
subsystem contains commands for managing the files in the
instrument and external USB stick.
The
LOG
subsystem contains the commands for managing the data logging of the
instrument.
................................................................................................165
DATA:DATA?
<>
Returns the logging file data of the selected file.
If manual trigger mode (trigger via TRIG function) is used, the logging function has to
be activated. Without activating the logging function in the manual trigger mode, the
instrument is not able to save a logging file internally or on the USB stick.
Data and File Management Commands
www.allice.de
Allice Messtechnik GmbH