Remote Control Commands
R&S
®
NGP800
162
User Manual 5601.5610.02 ─ 04
<numeric value>
Number of measurement values to be captured is set in the
range of 1 to 10000000.
MIN | MINimum
Minimum number of measurement values to be captured is set
at 1.
MAX | MAXimum
Maximum number of measurement values to be captured is set
at 10000000.
Parameters for setting and query:
<count>
MIN | MINimum | MAX | MAXimum
Returns the number of measurement values.
Example:
LOG:COUN MAX
LOG:COUN? MAX
-> 10000000
LOG:DATA?
Returns 12 sets of latest logging data with minimum logging interval (8 ms).
Depending on the models, the data is returned in the following format for a 2-channel
models: <Ch1_voltage>, <Ch1_current>,<Ch1_power>,<Ch2_voltage>, <Ch2_cur-
rent>,<Ch2_power>, <Ch1_voltage>, <Ch1_current>,<Ch1_power>, <Ch2_voltage>,
<Ch2_current>,<Ch2_power>...
Example:
2-channel models with 5 V set on Ch1 and 10V set on Ch2
LOG:DATA?
->
"5.000,0.0000,0.000,10.000,0.0000,0.000,5.000,0.0000,0.000,1
0.000,0.0000,0.000,5.000,0.0000,0.000,10.000,0.0000,0.000,5.
000,0.0000,0.000,10.000,0.0000,0.000,5.000,0.0000,0.000,10.0
00,0.0000,0.000,5.000,0.0000,0.000,10.000,0.0000,0.000,
5.000,0.0000,0.000,10.000,0.0000,0.000,5.000,0.0000,0.000,
10.000,0.0000,0.000,5.000,0.0000,0.000,10.000,0.0000,0.000,
5.000,0.0000,0.000,10.000,0.0000,0.000,5.000,0.0000,0.000,10
.000,0.0000,0.000,5.000,0.0000,0.000,10.000,0.0000,0.000"
Usage:
Query only
LOG:DURation
<Set new value>[, <Return min or max>]
LOG:DURation?
[<Return min or max>]
Sets or queries the duration of the data logging.
Setting parameters:
<Set new value>
<numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Duration of the data logging captured in the range of 0 s to
3.49*10^5 s.
MIN | MINimum
Minimum duration of the data logging captured at 0 s.