Instrument Tour
R&S
®
NGA100
20
Getting Started 5601.8902.02 ─ 03
4.1.3
Bottom Panel
The voltage selector is located at the bottom panel. On your first power-on con-
nection, you should see a yellow label sticker attached over the
. Before
peeling off the yellow label sticker, make sure that the correct fuse rating is used
for the mains voltage.
Figure 4-3: Bottom panel of R&S NGA100
12 = Voltage selector
13 = Voltage selector label
Voltage selector (12), Voltage selector label (13)
The voltage selector selects the mains voltage between 100 V, 115 V and 230 V.
See
To set the correct fuse rating, use a tool e.g. a flat screwdriver to move the selec-
tor according to the voltage selector label.
●
To select 100 V, slide both voltage selector to the left
●
To select 115 V, slide both voltage selector facing inward
●
To select 230 V, slide both voltage selector to the right
4.2
Switching On the Instrument
Before switching on the instrument, check that all the instructions in the “Basic
Safety Instruction” brochure and safety measures in previous sections are
observed.
Also, check if the value on the voltage selector corresponds to the mains voltage
(100 V, 115 V or 230 V).
Switching On the Instrument