Arbitrary and Function Generator
R&S AM300
E-1007.9922.12
6-182
Operating manual, 02/2005
6.2.7.2.1
Scaling the X Axis
Note
The
and
function keys are only available if the AUTO
function is switched off ( below).
Application
You can reduce or increase the scaling of the X axis and set the X axis
starting value. In so doing you define the visible time range.
In the AUTO function the R&S AM300 sets the X axis automatically. In the
case of periodic waveforms a period is then displayed across the whole
diagram area. After switching on the R&S AM300, the AUTO function is
active.
Scaling the X axis
automatically
1.
In the
submenu
press the
function key
.
The function key is highlighted and the X axis is automatically scaled.
Switching the
AUTO function off
2.
In the
submenu
press the
function key
.
The function key is
no longer
highlighted. The AUTO function is
switched off and the function keys for manually scaling the X axis are
available.
Scaling the X axis
manually
1.
In the
submenu
press the
function key
.
An entry field containing the currently applicable setting is displayed.
2.
Enter a new value ( 5-55).
The permissible entry range is:
8 ns
≤
T-Axis Scaling
≤
100000 s
in 1 ns steps
The X axis is re-scaled. The new time range defines the right and left
limits of the diagram area. The scale value of the vertical grid lines is
recalculated and displayed, e.g.
T-Axis Scaling = 10 ms: