R&S AM300
Arbitrary and Function Generator
Operating manual, 02/2005
6-105
E-1007.9922.12
6.2.1.4.4 EXPONENTIAL
Waveform
Application
You can set the start phase and the polarity for the
Exponential
waveform:
Selecting the
WAVE PARAM
submenu
1.
Select the
Exponential
waveform ( 6-72).
2.
In the
menu
press the
function key
.
The submenu is opened and the function keys [13] are assigned the
appropriate function.
Exit submenu
Set start phase
( 6-98)
Select polarity of waveform
( 6-106)
Note
The exponential function e
-x
is stored in RAM in the value range x = 0
to
x =
2*
π
(= 6.283). If you wish to produce a particular time constant, you can do
so by setting the frequency according to the formula f = 1/(2*
π
*T).
Individual adjustments can be performed with the aid of the Burst function.
START
PHASE
Norm
POLARITY
Inv