Arbitrary and Function Generator
R&S AM300
E-1007.9922.12
6-116
Operating manual, 02/2005
Setting the Modulation Frequency
Application
In the case of amplitude modulation, the output signal envelope changes as a
function of the frequency and amplitude of the modulation signal.
Setting the
AM frequency
1.
In the
submenu
press the
function key
.
The entry field for entering the frequency is displayed together with the
currently applicable setting. The default setting is "100 Hz".
2.
Enter a new value ( 5-55).
The permissible entry range is:
10 mHz
≤
Frequency
≤
100 kHz
in 10 µHz steps
The new setting is stored and then displayed in the parameter field if
modulation is switched on ( 6-117).
Setting the Modulation Depth
Application
The AM modulation depth m describes the ratio of the maximum to the
minimum amplitude A of the modulated signal.
min
max
min
max
A
A
A
A
m
+
−
=
Setting the
modulation depth
1.
In the
submenu
press the
function key
.
An entry field containing the currently applicable setting is displayed. The
default setting is "50 %".
2.
Enter a new value ( 5-55).
The permissible entry range for the modulation depth is.
0 %
≤
AM Depth
≤
100 %
in 1 % steps
The new setting is stored and then displayed in the parameter field if
modulation is switched on ( 6-117).
A
max
A
min
Summary of Contents for AM 300
Page 2: ...02 2005 ...