Arbitrary and Function Generator
R&S AM300
E-1007.9922.12
6-124
Operating manual, 02/2005
Setting the Modulation Frequency
Application
The frequency of the modulation signal is contained in the frequency of the
phase change in the output signal.
Setting the
PM frequency
1.
In the
submenu
press the
function key
.
The entry field for entering the frequency is displayed together with the
currently applicable setting. The default setting is "100 Hz".
2.
Enter a new value ( 5-55).
The permissible entry range is:
10 mHz
≤
PM Frequency
≤
100 kHz
in 10 µHz steps
The new setting is stored and then displayed in the parameter field if
modulation is switched on ( 6-125).
Setting the Phase Deviation
Application
The phase deviation ±
∆
ϕ
specifies the maximum deviation of the output
phase from the reference phase of the unmodulated carrier.
Setting the
phase deviation
1.
In the
submenu
press the
function key
.
An entry field containing the currently applicable setting is displayed. The
default setting is "0 deg".
2.
Enter a new value ( 5-55).
The permissible entry range for the phase deviation is:
0 deg
≤
PM Phase
≤
360 deg
in 0.01° steps
The new setting is stored and then displayed in the parameter field if
modulation is switched on ( 6-125).