Arbitrary and Function Generator
R&S AM300
E-1007.9922.12
6-160
Operating manual, 02/2005
6.2.4.4
Setting the Internal Trigger Generator Frequency
Application
You can enter the trigger frequency directly as a frequency value or indirectly
via the period.
Setting the
trigger frequency
1.
In the
menu
press the
function key
.
An entry field containing the currently applicable setting is displayed. The
default setting is "1 kHz".
2.
Enter a new value ( 5-55).
The permissible entry range is:
101 µHz
≤
Trigger Freq.
≤
2 MHz
in 1 µHz steps
3.
Press the
ENTER key
Setting the
trigger period
1.
In the
menu
press the
function key
.
The entry field containing the currently applicable setting is displayed.
The default setting is "1 ms".
2.
Enter a new value ( 5-55).
The permissible entry range is:
500 ns
≤
Trigger Period
≤
9901 s
in 10 ns steps
The new setting is stored.
Delivering a
trigger signal to
the Sync output
1.
Select the "Trigger" setting as the synchronization source ( 6-140).
2.
Select the "Internal" setting as the trigger mode ( 6-150).
3.
Switch the appropriate Sync output on ( 6-142).
Summary of Contents for AM 300
Page 2: ...02 2005 ...