Arbitrary and Function Generator
R&S AM300
E-1007.9922.12
6-94
Operating manual, 02/2005
Setting the Signal Amplitude
Application
You cannot enter a voltage directly for user-defined waveforms. The scaling
is used to define the response to voltage. This represents the peak-to-peak
voltage V
pp
when the full digital value range (+2
13
to
-2
13
) is being used.
The following relationship applies:
V
pp
= V’pp * scaling
This means that a digital value of -2
13
(+2
13
) in the waveform memory at a
scaling of 3 V/Full Scale yields an output voltage of 3 V
pp
at 50
Ω
.
Waveform memory
Output voltage with
scaling = 3 V/Full Scale
Setting the
scaling
1.
In the
submenu
press the
function key
.
An entry field containing the current setting is displayed. The default
setting is 2 V FS (Full Scale).
2.
Enter a new value ( 5-55).
The permissible entry range for the scale factor is:
1 mV
≤
scaling
≤
10 V
in 0.1 mV steps
Level display
The new setting is stored and then displayed in the diagram area. The
divisions on the level axis (grid lines) change in accordance with the set
scaling.
V’pp
+2
13
dig
-2
13
dig
Vpp