R&S ENV4200
Preparing for Use
User Manual 1107.2470.12 - 03
E-23
4.10 Connecting the EUT
The device under test is connected with the connecting cables to the front side of the
MC connector plug (L1, L2, L3, N). The maximum permissible continuous current
voltage is 200 A.
Suitable MC cable sockets for establishing connecting cables are included in the
equipment supplied.
The MC cable socket with the identification color blue and label N (neutral conductor)
is intended for connection to the MC high current connection with the 'N' identifier, see
. The MC cable socket with the identification color black and the labels L1, L2
and L3 ('phase' or 'line') are intended for connection to the MC high current
connections with the L1, L2 and L3 labels.
In order to mimic the influence of the user's hand while measuring the disturbance
voltage, usage of an "artificial hand" is required for EUTs that are normally held in the
hand. The artificial hand consists of a metal foil that must be connected to the RF
reference ground via an RC combination. The RC combination is a series connection
of a 220 pF capacitor C and a 510
Ω resistor R that is built into the R&S ENV4200. It
can be connected to the EUT via the 4 mm banana jack on the front (ARTIFICIAL
HAND).