Remote Control Commands
R&S
®
NGA100
102
User Manual 5601.8919.02 ─ 03
ARBitrary:STOP
This command stops the EasyArb waveform in channel 1 if EasyArb is running.
Usage:
Event
7.5.2
EasyRamp Commands
The
VOTage:RAMP
subsystem contains the commands for configuring the ramp func-
tion for the output channels.
[SOURce:]VOLTage:RAMP:DURation
<Time>
This command sets/queries the EasyRamp duration of the previous selected channel
in seconds. 0.01 s to 10 s in steps of 0.01 s.
Parameters:
<Time>
Range:
0.010 to 10
*RST:
0.010
Default unit: ms
[SOURce:]VOLTage:RAMP[:STATe]
<arg0>
This command defines/queries the EasyRamp state for the previous selected channel.
Parameters:
<arg0>
{OFF | ON}
ON
EasyRamp function is activated.
OFF
EasyRamp function is disabled.
7.5.3
DIO Commands
The
DIO
subsystem contains the commands for configuring a Digital I/O function for
the output channels.
TRIGger:DIRection:DIO<IO>
<arg0>
This command sets or queries the specified Digital I/O line to function as Trigger Input
or Trigger Output.
Suffix:
<IO>
.
1..4
Parameters:
<arg0>
OUTPut | INPut
Example:
TRIG:DIR:DIO2 OUT
Advanced Operating Commands