Remote Control Commands
R&S
®
NGA100
104
User Manual 5601.8919.02 ─ 03
INHibit
Output state remains off as long as the TRIG IN signal is activa-
ted.
ARBitrary
Starts the EasyArb function. If level signal is selected, the Easy-
Arb function is stopped when TRIG IN signal is deactivated.
*RST:
OFF
TRIGger:IN:SOURce:DIO<IO>
<arg0>
This command sets or queries the source under control of the trigger input of the speci-
fied Digital I/O line.
Suffix:
<IO>
.
1..4
Parameters:
<arg0>
CH1
Trigger input source from Ch 1.
CH2
Trigger input source from Ch 2.
*RST:
CH1
TRIGger:LOGic:DIO<IO>
<arg0>
This command sets or queries the trigger logic (Active High/Active Low) of the speci-
fied Digital I/O line.
Suffix:
<IO>
.
1..4
Parameters:
<arg0>
LOW | HIGH
*RST:
HIGH
TRIGger:MASTer[:STATe]
<arg0>
This command defines or queries the master enable state for the Digital I/O interface.
The state is disabled when any setting has been changed.
Parameters:
<arg0>
{OFF | ON}
ON
"Master Enable" state is activated for all DIO interfaces.
OFF
"Master Enable" state is deactivated for all DIO interfaces.
*RST:
OFF
Advanced Operating Commands