Remote Control Commands
R&S
®
NGA100
106
User Manual 5601.8919.02 ─ 03
TRIGger:OUT:CURRlevel:DIO<IO>
<Current>
This command sets or queries the threshold current level of the trigger output of the
specified Digital I/O line.
Triggers when output current level exceeds or equals to the set level.
Suffix:
<IO>
.
1..4
Parameters:
<Current>
*RST:
3.000
Default unit: A
TRIGger:OUT:SOURce:DIO<IO>
<arg0>
This command sets or queries the control source of the trigger output of the specified
Digital I/O line.
Suffix:
<IO>
.
1..4
Parameters:
<arg0>
CH1
Trigger output source from Ch 1.
CH2
Trigger output source from Ch 2.
*RST:
CH1
TRIGger:OUT:VOLTlevel:DIO<IO>
<Voltage>
This command sets or queries the threshold voltage level of the trigger output of the
specified Digital I/O line.
Triggers when output voltage level exceeds or equals to the set level.
Suffix:
<IO>
.
1..4
Parameters:
<Voltage>
*RST:
0.0
Default unit: V
TRIGger:RCL
<arg0>
This command recalls DIO settings from the respective memory locations M1-M5.
Parameters:
<arg0>
Range:
0..4
Example:
TRIG:RCL 1
Usage:
Setting only
Advanced Operating Commands