Important Notes
R&S
®
NGA100
12
User Manual 5601.8919.02 ─ 03
circuiting the fuse holder is prohibited. Resulting damages are not covered by the war-
ranty.
Safe operation
If the instrument is not in use, it must be switched off at the mains switch for safety rea-
sons.
3.5
Limits
The R&S NGA100 is equipped with a protective overload feature. The protective over-
load feature prevents damage to the instrument and is intended to protect against a
possible electrical shock. The maximum values for the instrument must not be excee-
ded. The protection limits are listed on the front panel of the R&S NGA100 to ensure
the safe operation of the instrument.
These protection limits must be adhered to:
Specification
Limits
Maximum output voltage
35 V module: 35.00 VDC
100 V module: 100.00 VDC
Maximum output current
35 V module: 6.0
100 V module: 2.0
Maximum voltage against earth
250 VDC
Maximum reverse voltage
35 V module: 36.00 VDC
100 V module: 102.00 VDC
Maximum inverse voltage
0.4 VDC
Maximum permitted current in case of inverse volt-
age
35 V module: 6.00 A
100 V module: 2.00 A
Power supply
100 VAC, 115 VAC or 230 VAC (tolerance
±
10 %)
Frequency
50 Hz / 60 Hz
Maximum power output
40 W (R&S NGA101, R&S NGA141)
80 W (R&S NGA102, R&S NGA142)
Limits