Getting Started
R&S
®
NGA100
25
User Manual 5601.8919.02 ─ 03
4.3.2
Setting the Output Voltage and Current Limit
To set the output voltage and current limit via Live-Mode:
1. Long press the rotary knob to enter into editing mode.
The voltage at channel 1 is highlighted in blue.
2. Move to the desired parameter via the arrow keys.
3. Rotate the knob to adjust its value.
Alternatively:
1. Press [Voltage] or [Current] key on the front panel.
2. For the two-channel R&S NGA100, press the desired channel key to activate the
respective voltage or current limit setting of that channel. The value on the respec-
tive channel becomes editable and is positioned by a blue cursor.
3. Press the [Left] or [Right] arrow key to move the cursor.
4. Press the [Up]/[Down] arrow key or turn the rotary knob to change the value.
The new value registers immediately.
4.3.3
Activating the Channels Output
The output voltages can be switched on or off regardless of the operating mode of the
instrument.
To activate the channel output, press the [Output] key on the front panel followed by
the desired channel key or vice versa.
For single channel models, press the [Output] key to activate the channel output.
Depending on the instrument operating mode, the display font color changes to green
in CV (constant voltage) mode and red in CC (constant current) mode.
4.3.4
Storing/Recalling of Instrument Settings
The instrument settings can be stored in the instrument memory by long pressing the
[Store Recall] key followed by the desired memory location key ([M1] to [M5]). The pre-
vious saved settings are overwritten.
To retrieve the settings, press [Store Recall] key and select the desired memory loca-
tion key ([M1] to [M5]).
Trying Out the Instrument