Operating Basics
R&S
®
NGA100
29
User Manual 5601.8919.02 ─ 03
Function
Description
DIO:xxxx
Digital Trigger I/O (Digital Trigger I/O option NGA-K103 must be installed).
The "xxxx" refers to I/O status for DIO1, DIO2, DIO3, DIO4.
If in use, the icon is highlighted in green.
If enabled, the icon is highlighted in white.
If disabled or option not installed, the icon is highlighted in gray.
Wireless LAN connection (WLAN option R&S NGA-K102 must be instal-
led).
If connected to access point, the icon is highlighted in green. It shows yel-
low when connecting to access point.
If disconnected or disabled, the icon is highlighted in gray.
TMC/VCP
USB connection with USB-VCP or USB-TMC setting.
If TMC or VCP is selected, the icon is highlighted in green.
Indicate buzzer state ("Disabled", "Fault Events", "Any Events").
If speaker is enabled, the icon is highlighted in green, else it shows gray.
Channel status bar
Function
Description
Channel number
Channel number indication.
Power
Display output power in watt.
SER/PAR
Indicator to show the series or parallel mode if chan-
nel fusion is enabled.
Available only for the two-channel models.
If enabled, the indicator displays in white.
SNS
Remote sense status indication.
If enabled, the indicator displays in green.
OVP / OPP / OTP
Indicator to show that the overload voltage, power or
temperature protection is triggered.
When triggered, the indicator shows flashing and
displays in red.
Fuse
Indicator to show the fuse status (on or off).
When triggered, the indicator shows flashing and
displays in red.
5.1.2
Channel Display Area
The R&S NGA100 displays two channels display area (Ch 1, Ch 2) for NGA102 and
NGA142 and a single channel display area for NGA101 and NGA141. The respective
channel settings and functions are displayed for each channel.
Display Overview