To avoid oscillation in "Auto" mode, the "Low" range to "High" range switching point is
at 210 mA and the "High" range to "Low" range switching point is at 170 mA.
Example:
The following shows the displays of current range set to "High" for Ch 1 and "Low" for
Ch 2.
If output current has exceeded 200 mA when current range is set "Low", the "OVLD"
information is displayed.
6.4
Remote Sense Function
The "Remote Sense" is a mechanism used to monitor and compensate the voltage
drops on the cables connected to the load.
1. Press [Menu] key.
2. Select "Output" > "Remote Sense" menu item to configure the remote sense mode.
The R&S NGA100 displays the "Remote Sense" dialog.
Remote Sense Function