Instrument Functions
R&S
®
NGA100
48
User Manual 5601.8919.02 ─ 03
4. Configure the "Fuse Linking" setting.
The R&S NGA100 provides the following options:
Example for "Ch 1", the available options are as follows:
● "Ch 2": "Ch 2" is linked to "Ch 1". If "Ch 1" fuse tripped, "Ch 2" is turned off.
● "Disabled": "Ch 1" is not linked to any other channels.
6.7
Tracking Function
Applicable only to NGA102 and NGA142 power supply models.
For power supplies with multiple channels, the channels can be linked such that
changes made on the selected tracked channel is applied to other channel.
1. Press [Track] key.
The R&S NGA100 activates the tracking function and the [Track] key illuminates.
2. Press the tracked channel to be tracked.
The selected tracked channel key ([Ch 1] or [Ch 2]) illuminates.
3. Press [Current] key if current values should be tracked.
All the channels are highlighted in blue and all the navigation controls illuminate.
Selected [Voltage] or [Current] key is also illuminated.
4. Set the desired voltage or current value.
Use the arrow keys or rotary knob to enter the value. All tracked channels voltage
or current are changed by the same amount.
6.8
Protection
The R&S NGA100 provides individual overvoltage protection (OVP) and overpower
protection (OPP) settings for each channel. With both function activated, the R&S
NGA100 protects itself and the load in case of voltage or power overloads.
Protection