The available options are as follows:
● "Enabled": The channel is turned off if the measured power exceeds OPP
value.
● "Disabled": Depending on models, output power is limited to maximum value of
40 W (NGA101, NGA102) or 80 W (NGA141, NGA142).
6. Set the desired power for OPP.
The R&S NGA100 turns off the respective channel when the measured power
exceeded the preset value.
6.9
Digital Trigger I/O
Digital I/O Pins Voltage Rating
Do not exceed the maximum voltage rating of the Digital I/O pins (5.25 V max) when
supplying voltages to the pins.
R&S NGA-K103 (order number: 5601.8425.99) option is required for the Digital Trigger
I/O.
The four data lines of the digital I/O interface are mutually independent and can be
used as trigger input or trigger output separately. See
●
Trigger input
The data lines of the digital I/O interface receive external trigger signal. The exter-
nal trigger signal triggers the selected channel (Ch 1, Ch 2) with the set response
when the trigger condition (Active High or Active Low) is met.
●
Trigger output
The data lines of the digital I/O interface output an "Active Low" or "Active High"
signal when the trigger condition of the selected channel is met.
Table 6-1: Trigger in parameters and conditions
Trigger in parameters
Trigger conditions
Description
Channel
Ch 1, Ch 2
Target output channel selected for trigger
response.
Response Trigger
Output On
Level
Selected channel [Output] is enabled when
the selected logic level is met.
Output Off
Level
Selected channel [Output] is disabled when
the selected logic level is met.
Output Toggle
Pulse
Selected channel [Output] is toggled when
the selected logic pulse is met.
Digital Trigger I/O