Remote Control Commands
R&S
®
NGA100
85
User Manual 5601.8919.02 ─ 03
[SOURce:]CURRent:RANGe
<arg0>
Sets or queries the current range of the selected channel.
Parameters:
<arg0>
AUTO | HIGH | LOW
AUTO
The measured output current is displayed in the following:
- "A" with 10 μA resolution for current above 200 mA
- "mA" with 1 μA resolution for current below 200 mA
HIGH
The measured output current is displayed in "A" with 10 μA res-
olution.
LOW
The measured output current is displayed in "mA" with 1 μA res-
olution.
7.3.5
Combined Setting of Voltage and Current Setting
The
APPLy
subsystem provides a command that enables you to set the current and
voltage of a channel in one step.
APPLy
<arg0>
This command defines or queries the voltage and current value of the selected chan-
nel.
Parameters:
<arg0>
{<Voltage> | DEF | MIN | MAX} [, {<Current> | DEF | MIN | MAX}]
<Voltage>
Set voltage in the range of 0 V to 35.05 V for NGA101 and
NGA102 models or 100.05 V for NGA141 and NGA142 models.
DEF
Set voltage to default.
MIN
Set voltage to minimum .
MAX
Set voltage to maximum.
<Current>
Set current in the range of 1 mA to 6.05 A for NGA101 and
NGA102 models or 1 mA to 2.03 A for NGA141 and NGA142
models.
DEF
Set current to default.
MIN
Set current to minimum.
MAX
Set current to maximum.
Configuration Commands