Remote Control Commands
R&S
®
NGA100
91
User Manual 5601.8919.02 ─ 03
FUSE:LINK 2
// queries the combined overcurrent protection of the selected channel
FUSE:LINK? 2
// response 1
// ************************************************
// Unlink linked overcurrent protection
// ************************************************
FUSE:UNLink 2
// queries the combined overcurrent protection of the selected channel
FUSE:LINK? 2
// response 0
FUSE:DELay
<Delay>
This command defines and queries the fuse delay for the previous selected channel.
Parameters:
<Delay>
{<Delay> | MIN | MAX}
<delay>
0 s to 10 s (adjustable in 0.010 s steps e.g. FUSE:DEL 0.010 =
0.01 s)
MIN
Set fuse delay to minimum.
MAX
Set fuse delay to maximum.
Default unit: ms
Example:
INST OUT1
FUSE:DEL 0.050
Set fuse delay of channel 1 to 50 ms.
FUSE:DEL?
-> 0.05
Return the fuse delay time of the selected channel.
FUSE:LINK
<arg0>
This command defines or queries the channel fuses (fuse linking) for the previous
selected channel.
Parameters:
<arg0>
1
channel Ch 1
2
channel Ch 2
Example:
INST OUT1
FUSE:LINK 2
Fuse Ch 1 is linked with Ch 2
FUSE:LINK? 2
-> 1
If the fuse of Ch 1 is linked with Ch 2, a "1" is returned.
Configuration Commands