Remote Control Commands
R&S
®
NGA100
96
User Manual 5601.8919.02 ─ 03
Example:
INST OUT1
VOLT:PROT ON
OVP of Ch 1 is activated.
VOLT:PROT?
-> 1
Return the OVP state of the selected channel.
[SOURce:]VOLTage:PROTection:LEVel
<Voltage>
This command defines or queries the OVP value of the previous selected channel.
Parameters:
<Voltage>
{<Voltage> | MIN | MAX}
<Voltage>
Set the OVP value in the range of 0 mV up to 36 V for NGA101
and NGA102 models or 0 mV up to 102 V for NGA141 and
NGA142 models (adjustable in 100 mV steps):
MIN
Set the OVP value to minimum.
MAX
Set the OVP value to maximum.
Default unit: V
Example:
INST OUT1
VOLT:PROT:LEV 5
Set the OVP value of Ch 1 to 5 V.
VOLT:PROT:LEV?
Return the selected channel OVP value.
[SOURce:]VOLTage:PROTection:CLEar
This command resets the OVP state of the selected channel. If the OVP has tripped
the OVP message on the display will be cleared for the selected channel.
Usage:
Event
7.3.9
OPP Setting
The
POWer:PROTection
subsystem contains the commands for setting the over-
power protection parameters for the output channels. The default unit is W.
Configuration Commands