Getting Started
R&S
®
NGA100
18
User Manual 5601.8919.02 ─ 03
4.1.4.1
Bench Operation
On a benchtop, the R&S NGA100 can either lie flat or stand on its feet. As shown in
, feet on the bottom can be folded out to set the instrument in an inclined
position.
Figure 4-1: Operating positions
Positioning of instrument
The instrument must be positioned in a manner that allows the user to disconnect the
unit from the mains at any time and without restrictions.
Risk of injury if feet are folded out
The feet can fold in if they are not folded out completely or if the instrument is shifted.
Collapsing feet can cause injury or damage the instrument.
●
Fold the feet completely in or out to ensure stability of the instrument. Never shift
the instrument when the feet are folded out.
●
When the feet are folded out, do not work under the instrument or place anything
underneath.
●
The feet can break if they are overloaded. The overall load on the folded-out feet
must not exceed 250 N.
4.1.4.2
Rack Mounting
The instrument can be installed in a 19" rack using the rack adapter R&S HZN96 (P/N
3638.7813.02). Proceed according to the installation instructions supplied with the rack
adapter.
Putting into Operation