Instrument Functions
R&S
®
NGA100
51
User Manual 5601.8919.02 ─ 03
Trigger in parameters
Trigger conditions
Description
Output Inhibit
Level
Selected channel [Output] is inhibited when
the selected logic level is met.
When the selected channel [Output] is put to
inhibit state, manual or remote operation on
selected channel [Output] is no longer possi-
ble .
To reverse the inhibit state, remove the
source of the trigger signal. You can either
disable the affected DIO interface or remove
the source from the affected DIO interface at
the rear panel.
Start EasyArb
(Only for Ch 1)
Level
Activates the EasyArb function on channel 1
when the selected logic level is met. The
EasyArb is deactivated when the level
changes (i.e from active high to active low)
See
Pulse
Activates the EasyArb function on channel 1
when the selected pulse is met. See
Logic
Active High
Set the logic level of the trigger in signal.
Active Low
Table 6-2: Trigger out parameters and conditions
Trigger out parameters Trigger conditions
Description
Channel
Ch 1, Ch 2
Output channel selected to monitor for trig-
ger condition.
Condition Trigger
Output On
Output the selected logic level when the
[Output] is turned on at the selected chan-
nel.
Output Off
Output the selected logic level when the
[Output] is turned off at the selected chan-
nel.
Fuse Tripped
Output the selected logic level when a fuse
tripped event occurs on the selected chan-
nel.
See
Chapter 6.6, "Activating Fuse"
CC Mode
Output the selected logic level when the
selected channel operates in the CC mode.
See
Voltage Level
>= "set value"
Output the selected logic level when the
voltage level of the selected channel is
greater or equal to the set level.
Current Level
>= "set value"
Output the selected logic level when the
current level of the selected channel is
greater or equal to the set level.
Digital Trigger I/O