Remote Control Commands
R&S
®
NGL200/NGM200
101
User Manual 1178.8736.02 ─ 06
MAX
Max value for lower safety limit.
Range:
For up to 6V: 0.001E+00 to 3.010E+00. For above
6V: 0.001E+00 to 6.010E+00
Increment: 0.001
*RST:
0.001E+00
Example:
See
Example "Configuring the current output"
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]
<current>
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]?
Sets or queries the upper safety limit for current.
Setting parameters:
<current>
numeric | MIN | MAX
numeric
Numeric value for upper safety limit.
MIN
Min value for upper safety limit.
MAX
Max value for upper safety limit.
Range:
For up to 6V: 0.001E+00 to 3.010E+00. For above
6V: 0.001E+00 to 6.010E+00
Increment: 0.001
*RST:
6.010E+00
Example:
See
Example "Configuring the current output"
7.5.3
Voltage Setting
The
SOURce:VOLTage
subsystem contains the commands for setting the voltage of
the output channels. The default unit is V.
Configuration Commands