Remote Control Commands
R&S
®
NGL200/NGM200
126
User Manual 1178.8736.02 ─ 06
OFF | 0
OPP is deactivated
ON | 1
OPP is activated
Example:
See
Example "Configuring the overpower protection"
[SOURce:]POWer:PROTection:CLEar
Resets the OPP state of the selected channel. If an OPP event has occurred before,
the reset also erases the message on the display.
Example:
See
Example "Configuring the overpower protection"
Usage:
Event
[SOURce:]POWer:PROTection:LEVel
<power>
[SOURce:]POWer:PROTection:LEVel?
Sets or queries the overvoltage protection value of the selected channel.
Parameters:
<power>
numeric | MIN | MAX | DEFault
numeric
Numeric value of the power protection level in watts.
MINimum
Minimum value of the power protection level at 0.00 W.
MAXimum
Maximum value of the power protection level at 60.00 W.
Range:
0.00 to 60.00
*RST:
60.00
Default unit: W
[SOURce:]POWer:PROTection:TRIPped?
Queries the OPP state of the selected channel.
Example:
See
Example "Configuring the overvoltage protection"
POW:PROT:TRIP?
Response 1, the OPP is tripped.
Response 0, the OPP is not tripped.
Usage:
Query only
Configuration Commands