Remote Control Commands
R&S
®
NGL200/NGM200
135
User Manual 1178.8736.02 ─ 06
ARBitrary:TRIGgered:MODE
<mode>
ARBitrary:TRIGgered:MODE?
Sets or queries the arbitrary trigger mode of the previous selected channel.
Parameters:
<mode>
SINGle | RUN
SINGle
A trigger event starts only with one arbitrary sequence.
RUN
A trigger event starts the whole arbitrary sequences (with all rep-
etitions).
ARBitrary:TRIGgered[:STATe]
<state>
ARBitrary:TRIGgered[:STATe]?
Sets or queries the arbitrary trigger mode.
Parameters:
<state>
0 | 1
0
OFF - Trigger input is deactivated.
1
ON - Trigger input is activated.
Example:
INST OUT1
ARB:TRIG ON
ARB:TRIG?
-> 1
Activates the Ch1 trigger mode for arbitrary function.
7.7.2
Ramp
The
VOTage:RAMP
subsystem contains the commands for configuring the ramp func-
tion for the output channels.
................................................................................. 135
[SOURce:]VOLTage:RAMP:DURation
.............................................................................. 136
[SOURce:]VOLTage:RAMP[:STATe]
<state>
[SOURce:]VOLTage:RAMP[:STATe]?
Sets or queries the state of ramp function for the previous selected channel.
Parameters:
<state>
0 | 1
0
EasyRamp function is deactivated.
1
EasyRamp function is activated.
Advanced Operating Commands