Remote Control Commands
R&S
®
NGL200/NGM200
140
User Manual 1178.8736.02 ─ 06
Example:
INST:NSEL 1
BATT:SIMU:RES?
Queries channel 1 of battery simulator internal resistance.
Usage:
Query only
BATTery:SIMulator:SOC
<arg0>
Sets or queries the state of charge (SoC) of the battery simulator.
Parameters:
<arg0>
Sets SoC values.
Example:
INST:NSEL 1
BATT1SIM:SOC 20
Sets the channel 1 Soc to 20 %.
BATTery:SIMulator:TVOLtage?
Queries the terminal voltage (Vt) of the specific channel.
Example:
INST:NSEL 1
BATT:SIM:TVOL?
Returns channel 1 terminal voltage of battery simulator.
Usage:
Query only
BATTery:SIMulator:VOC?
Queries the open circuit voltage (Voc) of the specific channel.
Example:
INST:NSEL 1
BATT:SIM:VOC?
Returns the channel 1 open circuit voltage of battery simulator.
Usage:
Query only
BATTery:SIMulator:VOC:EMPTy?
Queries the open circuit voltage (Voc) for empty SoC, i.e SoC = 0 %.
Example:
INST:NSEL 1
BATT:SIM:VOC:EMPT?
Returns the channel 1 Voc of an empty charge battery.
Usage:
Query only
BATTery:SIMulator:VOC:FULL?
Queries the open circuit voltage (Voc) for full SoC, i.e SoC = 100 %.
Advanced Operating Commands