Remote Control Commands
R&S
®
NGL200/NGM200
146
User Manual 1178.8736.02 ─ 06
Setting parameters:
<filepath>
Filepath of the file.
Example:
DATA:DEL
"/int/logging/log-20201203T095013.965.csv"
Deletes internal logging file 'log-20201203T095013.965.csv'
Usage:
Setting only
DATA:LIST?
Queries all files in internal memory ('/int/') and external memory ('/USB').
Example:
DATA:LIST?
-> "/USB1/NGL201/logging/
log-20201203T101025.829.csv", "/int/arb/
newWaveform.csv","/int/logging/log-20201203T101129.818.csv"
Usage:
Query only
DATA:POINts?
<filepath>
Queries the number of measurements from the selected logging file.
If manual trigger mode (trigger via TRIG function) is used, the logging function has to
be activated. Without activating the logging function in the manual trigger mode, the
instrument is not able to save a logging file internally or on the USB stick.
Parameters:
<filepath>
Filepath of the logging file data.
Example:
DATA:POIN?
"/USB1/NGL201/logging/log-20201203T101025.829.csv"
-> 5
Returns 5 log files counts from "/USB1/NGL201/logging/
log-20201203T101025.829.csv".
Usage:
Query only
FLOG:DATA?
Queries FastLog data as a block.
The block is returned in the binary format starting in the sequence of voltage followed
by current measurements, i.e. V, I, V, I, ....
The R&S
NGL/NGM accepts the line message EOI and/or the ASCII character NL
(0Ah) as an indication that data transmission has been completed
The binary data stream must be concluded with EOI or NL or EOI followed by NL. If the
data stream is not concluded with either EOI or NL, the R&S
NGL/NGM will wait for
additional data. In the case of a binary data transmission, the R&S
NGL/NGM ignores
the bit combination NL (0Ah) within the data stream.
Data and File Management Commands