Operation
Power Supply NGMO
4.00 / 07-2004
4-11
4.2.3.4 Menu
Items
The menu includes a selection of items from
A01
or
B01
to
A38
or
B38
:
(The indicated channel
A
or
B
depends on which channel is selected.)
A/B01:
Current range
[AUTO] 5A .5A 5mA
This function selects the range for static current measure-
ment (see main screen) and also selects the current range
for signal analysis. The analysis mode is only available in
the ranges .SA and SA.
A/B02:
Static Measure
Interval (2-200 ms)
This setting specifies the repetition time between two static
current and DVM measurements. Any value from 2 to 200
ms can be selected.
A/B03:
Static Measure
Average Count (1-10)
This setting specifies the average count for static current
and DVM measurements. Values from 1 to 10 can be
selected. The total measurement time for static measure-
ments is the product of “Static Measure Interval” multiplied
by “Static Measure Average Count”.
A/B04:
Output Impedance
(0-1.00 Ohms)
This menu item lets you set the output impedance from
0.00 Ohms to 1.00 Ohms im 10 mOhms steps.
A/B05:
Output Bandwidth
high low
This menu item lets you select one of two different output
regulation characteristics. When in “High” mode the
NGMO2 has the fasted regulation speed available (see
specifications for further information), but this may result in
an unstable load regulation for inductive loads.
Summary of Contents for NGMO1
Page 8: ...Power Supply NGMO Tables 4 Version 4 00 07 2004 ...
Page 14: ...Power Supply NGMO Safety 2 2 4 00 07 2004 ...
Page 51: ...Operation Power Supply NGMO 4 00 07 2004 4 29 Fig 4 7 Status model structure ...
Page 52: ...Power Supply NGMO Operation 4 30 4 00 07 2004 ...
Page 84: ...Power Supply NGMO Ordering Information 6 2 4 00 07 2004 ...