Unit Description
Power Supply NGMO
A serial interface RS232 and an IEEE-488-2 connection accommodate
remote control. Apart from that the unit contains a Control Interface with
control lines and relay drivers.
4.2.2
Keyboard / Display
The NGMO control unit consists of the keyboard with rotary transducer
and the detached LC display.
The display is connected to the keyboard via a cable for data and pow-
er supply and another two-pole cable for the LED background illumina-
tion. From the keyboard a common ribbon cable runs to the processor
board. The control processor on the processor board reads out keys
and rotary transducer and presents the up-to-date data on the display.
4.2.3
Analog Boards, CHN ”A“ and ”B“
The two analog boards Channel A and B (NGMO2 only) are electrically
of identical design and offer the following functions beside the typical
functions of a mains supply unit.
•
Adjustable output impedance
•
fast Push-Pull output stage with reversible bandwidth
•
electronic load (current sink)
•
high measuring and adjustment accuracy
•
quick sampling of output currents and of DVM-voltage with a sam-
pling frequency of up to 100kHz (transient measurement)
•
Data storage with 5000 storage locations
•
Analysis of recorded current and voltage characteristics
In addition the following safety functions:
•
Detection of open sensing lines
•
Overvoltage protection (cutoff in case of too high output voltage,
limit adjustable)
•
Overcurrent protection (cutoff or limitation to the set current value)
Communication between processor board and one or both analog
boards takes place via a serial interface, which is galvanically separat-
ed by an optical isolator. The analog boards are equipped with a signal
3.00 / 05-2014
4-3
Summary of Contents for NGMO1
Page 12: ...Power Supply NGMO User Information 1 4 3 00 05 2014 ...
Page 14: ...Power Supply NGMO Customer Service 2 2 3 00 05 2014 ...
Page 16: ...Power Supply NGMO Safety 3 2 3 00 05 2014 ...
Page 22: ...Power Supply NGMO Unit Description 4 6 3 00 05 2014 ...
Page 42: ...Power Supply NGMO Calibration 7 12 3 00 05 2014 ...
Page 72: ...Power Supply NGMO Interfaces 10 4 3 00 05 2014 ...
Page 74: ...Power Supply NGMO Bill of Materials 11 2 3 00 05 2014 ...