Power Supply NGMO
Calibration
The display must show the following:
Firmw.:x.yy CAL:zzzz
Lst CAL on: yy.mm.dd
•
x.yy is the firmware version currently loaded
in the NGMO.
x is the major version number and
yy is the minor version number.
•
zzzz: is the calibration counter. This counter
is incremented by one every time a calibra-
tion has been successfully completed.
Each channel has its own calibration coun-
ter.
•
yy.mm.dd is the date of the last calibration
for the selected channel.
yy specifies the calibration year,
mm is the calibration month,
dd is the calibration day.
8. Press ENTER.
The display must show the following:
Enter calibration
entry ID:
NOTE:
The calibration entry ID is identical for both channels.
The calibration entry ID is a number not longer than 8 digits which
acts as a password when you enter the real calibration functions. The
default factory ID is: 0.
After the correct number on the numerical keys has been entered the
NGMO enters the real calibration functions.
7-4
3.00 / 05-2014