Calibration
Power Supply NGMO
16. Enter the resistance value.
Use either the variation knob or numerical entry keys to enter the
exact resistor value.
17. Press ENTER.
NOTE:
When using the calibration set-up specified in section 7.2, the
following display may be neglected.
The display must show the following:
Connect known 33 Ohm
Resistor to out.term
18. Press ENTER.
The display must show the following:
Enter exact Resistor
value: 3x.xxx Ohms
NOTE:
The values of the resistors installed in the calibration adapter
can be taken from the calibration certificate supplied with the
calibration adapter.
19. Enter the resistance value.
Use either the variation knob or numerical entry keys to enter the
exact resistor value.
20. Press ENTER.
NOTE:
When using the calibration set-up specified in section 7.2, the
following display may be neglected.
3.00 / 05-2014
7-7
Summary of Contents for NGMO1
Page 12: ...Power Supply NGMO User Information 1 4 3 00 05 2014 ...
Page 14: ...Power Supply NGMO Customer Service 2 2 3 00 05 2014 ...
Page 16: ...Power Supply NGMO Safety 3 2 3 00 05 2014 ...
Page 22: ...Power Supply NGMO Unit Description 4 6 3 00 05 2014 ...
Page 42: ...Power Supply NGMO Calibration 7 12 3 00 05 2014 ...
Page 72: ...Power Supply NGMO Interfaces 10 4 3 00 05 2014 ...
Page 74: ...Power Supply NGMO Bill of Materials 11 2 3 00 05 2014 ...