Calibration
Power Supply NGMO
After a short time, the NGMO displays the following message:
Really set cal.entry
ID to 0. (ENTER=YES)
Press ENTER.
If you press the ENTER key, the current calibration entry ID will be
overwritten with a zero. Any other key aborts this function.
7. Remove the calibration entry ID jumper.
8. Switch off the NGMO and remove the power cord.
9. Replace the main shielding on the device and mount the four feet
to the rear of the device. (see section 9.2.5)
3.00 / 05-2014
7-11