When a DPDT relay is
not activated
(no voltage applied to the relay) and after a
reset, the terminals
1 + 2
and
3 + 4
are connected. The user interface (control
screen / "WebGUI") labels this state as "Pos = 0".
In the remote control command
, this deactivated state of the relay
is represented by "
00
".
●
When the relay is
activated
(continuous voltage applied to the relay), the terminals
1 + 3
and
2 + 4
are connected. The user interface (control screen / "WebGUI")
labels this state as "Pos = 1".
In the remote control command
, this activated state of the relay is
represented by "
01
".
The relays are mounted directly into the module's panel and soldered to the module's
printed board, to avoid relay cabling and allow a compact module size.
The switch unit provides power supply and control via a single module bus cable to a
module bus connector on the module's printed board. The module's on-board memory
stores its configuration data.
Figure 3-130: Pinout and block diagram of RF switch module R&S
OSP-B136
Figure 3-131: Representation of the module on the control screen
For operation, see
R&S
OSP-B136 RF Switch Module