Getting Started
R&S
®
OSP
31
User Manual 1178.7100.02 ─ 05
2.2.2.4
Rear Panel of the R&S
OSP-B200S2 Satellite Unit
For a comprehensive description of this unit, also refer to the
R&S
OSP-B200R/
B200S2 Satellite System user manual
, available for download at
The rear panel of the R&S
OSP-B200S2 features various connectors and LEDs.
You cannot insert any modules into the rear panel of the R&S
OSP-B200S2.
Figure 2-12: Rear view of the R&S
OSP-B200S2 (2HU)
1 = DC power supply connector
2 = Fiber-optic link (FOL) connector for optical remote control
3 = Wired link connector for electrical remote control
4 = Status LEDs for indicating [Power] and [Overheat]
5 = Status LED for indicating [Link / Busy]
Wired Link versus Fiber-Optic Link
The link connectors (labeled 2 and 3 in
) allow choosing either an electrical
or an optical control connection.
●
Wired link
For remote operation of the satellite unit R&S
OSP-B200S2 across distances up to
10
m, use the D-Sub cable R&S
OSP-Z200x.
Connect this cable
exclusively
to the remote control interface module
in your base switch unit. Connecting it to any other device or module can
harm your equipment.
The wired link cable supplies power from the base switch unit to the satellite.
●
Fiber-optic link
For remote operation of the satellite unit R&S
OSP-B200S2 across larger distan-
ces up to 20
m, or inside a shielded room, use the fiber-optic cable R&S
OSP-
Z201x or R&S
OSP-Z202x.
As the fiber-optic link cable does not supply power to the satellite unit, you must
also use the 28
V
DC power supply R&S
OSP-B200P.