RC Commands
R&S
®
OSP
267
User Manual 1178.7100.02 ─ 05
......................................................................................279
.............................................................................................279
..............................................................................................280
CONFigure:FRAMe:IMPort:CATalog?
...............................................................................280
..................................................................................281
CONFigure:FRAMe:IMPort:DELete:ALL
........................................................................... 281
CONFigure:POWerup:RESet
<state>
Sets or queries the Power Up reset condition of switch modules with
latching
relays.
This setting determines, how latching relays behave after booting the switch unit.
Note that this command does
NOT reset
the module
, which is designed for
controlling
external
latching relays.
The following rules apply for identifying positions of
latching SPDT
relays:
●
On the relays' front plates, the positions are labeled as
"2"
and
"1"
, with "2" being
the default position
●
In the graphical user interface (GUI), the positions are
"0"
and
"1"
, with "0" being
the default position
●
In a remote control command (SCPI), the positions are
"00"
and
"01"
, with "00"
being the default value
Hence, if you take the front-plate port labels of a latching SPDT relay, subtract "1" to
get the position values that the software uses for this relay.
The query returns the current reset condition.
Setting parameters:
<state>
1 | ON
At Power Up, the switch unit handles latching relays as follows:
It sets all latching
SPDT
relays to the default ports labeled
"2"
,
which are represented in the software by position
"0"
.
It sets all latching
SPxT
relays to the open state.
0 | OFF
At Power Up, the switch unit leaves all latching relays keep their
previous state.
Example:
CONF:POW:RES?
Response:
1
Here, the reset condition is "
1
", equivalent with
being enabled.
Manual operation:
See
CONFigure:POWerup:PATH
<path name>
Sets or queries the switch-on action that determines, which path (if any) is switched
after booting the instrument.
This RC command is equivalent to the
configuration
dialog of the user interface.
CONFigure Commands