RC Commands
R&S
®
OSP
282
User Manual 1178.7100.02 ─ 05
8.6
TRIGger Commands
The following commands provide remote control of your switch unit's trigger functional-
ity, described in
Some modules are not triggerable, for example R&S
OSP-B102L and R&S
OSP-
B112UL, due to their multi-position latching relays. For detailed information on compati-
bility of your modules with the trigger functionality, refer to the
dia-
log or to the data sheet.
...............................................................................................283
...........................................................................................284
...............................................................................................285
........................................................................................... 286
TRIGger:SEQuence:DEFine:LENGth
............................................................................... 287
TRIGger:SEQuence:DEFine[:ENTRy]
.............................................................................. 287
.....................................................................................287
.........................................................................................288
TRIGger:TYPE
<mode>
Selects or queries the trigger type.
Parameters:
<mode>
SINGle
Selects the trigger type "Single".
TOGGle
Selects the trigger type "Toggle A-B".
SEQuenced
Selects the trigger type "Sequenced".
ADDRessed
Selects the trigger type "Addressed".
Example:
Setting:
TRIG:TYPE SEQ
Query:
TRIG:TYPE?
Reply:
SEQ
TRIGger Commands