RC Commands
R&S
®
OSP
288
User Manual 1178.7100.02 ─ 05
Parameters:
<numeric_value>
Specifies the voltage level. Factory default is 2.5
V.
Range:
0.5 to 4.5
Increment: 0.01
Default unit: V|mV
Example:
Setting:
TRIG:SIGN:LEV 0.5V
Query:
TRIG:SIGN:LEV?
Reply:
5.000000E-01
Setting:
TRIG:SIGN:LEV 4.5V
Query:
TRIG:SIGN:LEV?
Reply:
4.00
Manual operation:
See
TRIGger:SIGNal:SLOPe
<slope>
Sets or queries, which kind of threshold transition is interpreted as a trigger event. For
the threshold level, see
Parameters:
<slope>
POSitive
Defines interpreting a positive (low-to-high) transition of the
threshold as a trigger event.
NEGative
Defines interpreting a negative (high-to-low) transition of the
threshold as a trigger event.
BOTH
Defines interpreting a positive or a negative transition of the
threshold as a trigger event.
Equivalent with "Any" in the user interface ("WebGUI").
Example:
Setting:
TRIG:SIGN:SLOP POS
Query:
TRIG:SIGN:SLOP?
Reply:
POS
Manual operation:
See
TRIGger:SIGNal:TERMination
<term>
Sets or queries the type of termination of the trigger signal cable.
Parameters:
<term>
50OHM
Sets the termination to "50 Ohm".
Also accepted parameter syntax: "
50
" or "
50 ohm
" (a blank and
all capital/small letters are ignored).
TRIGger Commands