Remote Control Commands
R&S
®
NGL200/NGM200
133
User Manual 1178.8736.02 ─ 09
[SOURce:]CURRent:PROTection:TRIPped?
[<Channel list>]
Queries the OCP state of the selected channel.
Query parameters:
<Channel list>
<list>
Example:
CURR:PROT:TRIP?
Response 1, the OCP is tripped.
Response 0, the OCP is not tripped.
Example:
CURR:PROT:TRIP? (@1)
Queries OCP state at channel 1.
Example:
See
Example "Configuring the overcurrent protection"
Usage:
Query only
[SOURce:]CURRent:PROTection:UNLink
<arg0>
Unlink fuse linking from the other channel (Ch1 or Ch2).
Parameters for setting and query:
<arg0>
1 | 2
Example:
CURR:PROT:UNL 1
Fuse linking is unlinked from channel 1
Example:
See
Example "Configuring the overcurrent protection"
Usage:
Setting only
[SOURce:]CURRent:PROTection[:STATe]
<arg0>[, <Channel list>]
[SOURce:]CURRent:PROTection[:STATe]?
[<Channel list>]
Sets or queries the OCP state.
Parameters:
<arg0>
1
Activates the OCP state.
0
deactivates the OCP state.
Parameters for setting and query:
<Channel list>
<list>
Example:
CURR:PROT 1
Activates the OCP.
Example:
CURR:PROT? (@1)
Queries OCP state at channel 1.
Example:
See
Example "Configuring the overcurrent protection"
Configuration Commands