Remote Control Commands
R&S
®
NGL200/NGM200
135
User Manual 1178.8736.02 ─ 09
FUSE:LINK
<arg0>
FUSE:LINK?
<arg0>
Sets or queries the fuses of several serlected channels (fuse linking).
Parameters for setting and query:
<arg0>
1 | 2
Example:
INST OUT1;:FUSE:LINK 2
Channel 2 is linked with channel 1
INST OUT1;:FUSE:LINK?2
Return 1 if channel 1 is linked with channel 2
Example:
Example "Configuring the over-
FUSE:TRIPped?
[<Channel list>]
Queries the status if fuse has tripped in the selected channel.
Parameters:
<Channel list>
<list>
Example:
FUSE:TRIP? (@1)
Queries fuse tripped status at channel 1.
Example:
For alternative command, see
Example "Configuring the over-
Usage:
Query only
FUSE:UNLink
<channel>
Dissolves linked fuses.
Parameters:
<channel>
1 | 2
Example:
For alternative command, see
Example "Configuring the over-
Usage:
Setting only
FUSE[:STATe]
<state>
FUSE[:STATe]?
[<Channel list>]
Sets or queries the fuse function in the selected channel.
Parameters:
<state>
1
Fuse function is activated.
0
Fuse function is not activated.
Configuration Commands