Remote Control Commands
R&S
®
NGL200/NGM200
139
User Manual 1178.8736.02 ─ 09
MAX | MAXimum
Maximum value for the overvoltage protection value at 20.05 V.
DEF | DEFault
Default value of the overvoltage protection level at 20.05 V.
Range:
0.001 to 20.05
*RST:
20.05
Default unit: V
Parameters for setting and query:
<Channel list>
<list>
Example:
VOLT:PROT:LEV? (@1)
Queries overvoltage protection value at channel 1.
Example:
See
Example "Configuring the overvoltage protection"
[SOURce:]VOLTage:PROTection:TRIPped?
[<Channel list>]
Queries the OVP state of the selected channel.
Parameters:
<Channel list>
<list>
Example:
VOLT:PROT:TRIP?
Response 1, the OVP is tripped.
Response 0, the OVP is not tripped.
Example:
VOLT:PROT:TRIP? (@1)
Queries OVP state at channel 1.
Example:
See
Example "Configuring the overvoltage protection"
Usage:
Query only
7.5.11
OPP Setting
The
POWer:PROTection
subsystem contains the commands for setting the over-
power protection parameters for the output channels. The default unit is W.
Configuration Commands