Remote Control Commands
R&S
®
NGL200/NGM200
146
User Manual 1178.8736.02 ─ 09
Example:
MEAS:CURR:DC:MIN? (@1)
Queries the minimum measured output current at channel 1.
Usage:
Query only
MEASure[:SCALar]:CURRent[:DC]:STATistic?
[<Channel list>]
Queries the current statistics of the selected channel.
Parameters:
<Channel list>
<list>
Example:
INST:NSEL 1
MEAS:CURR:DC:STAT?
Queries the current statistics at channel 1.
Example:
MEAS:CURR:DC:STAT? (@1)
Queries the current statistics at channel 1.
Usage:
Query only
MEASure[:SCALar]:POWer?
[<Channel list>]
Queries the currently emitted power of the selected channel.
Parameters:
<Channel list>
<list>
Example:
MEAS:POW?
-> 3.00E+00
Example:
MEAS:POW? (@1)
Queries the currently supplied power at channel 1.
Usage:
Query only
MEASure[:SCALar]:POWer:AVG?
[<Channel list>]
MEASure[:SCALar][:POWer]:AVG?
Queries the average measured output power.
Example:
MEAS:POW:AVG? (@1)
Queries the average measured output power at channel 1.
Usage:
Query only
MEASure[:SCALar]:POWer:MAX?
[<Channel list>]
MEASure[:SCALar][:POWer]:MAX?
Queries the maximum measured output power.
Example:
MEAS:POW:MAX? (@1)
Queries the maximum measured output power at channel 1.
Usage:
Query only
Measurement Commands