Remote Control Commands
R&S
®
NGL200/NGM200
163
User Manual 1178.8736.02 ─ 09
BATTery:SIMulator[:ENABle]
<arg0>[, <Channel list>]
BATTery:SIMulator[:ENABle]?
[<Channel list>]
Sets or queries the specific channel of the battery simulator state.
Parameters:
<arg0>
1
Enables the battery simulator state.
0
Disables the battery simulator state.
Parameters for setting and query:
<Channel list>
<list>
Example:
INST:NSEL 1
BATT:SIM 1
Enables the battery simulator state for channel 1.
Example:
BATT:SIM 1, (@1)
Enables the battery simulator state at channel 1.
BATTery:STATus?
[<Channel list>]
Queries the status of the battery (idle, charging or discharging).
Parameters:
<Channel list>
<list>
Example:
INST:NSEL 1
BATT:STAT?
Returns the status of the battery simulator from channel 1.
Example:
BATT:STAT? (@1)
Queries the status of battery simulator at channel 1.
Usage:
Query only
BATTery:MODel:CURRent:LIMit:EOC
<arg0>
BATTery:MODel:CURRent:LIMit:EOC?
Sets or queries the current limit of the battery model at end-of-charge.
Parameters:
<arg0>
Sets the current limit of the battery model at end-of-charge.
Example:
INST:NSEL 1
BATT:MODel:CURR:LIM:EOC?
Returns current limit of the battery model at end-of-charge for
channel 1.
Advanced Operating Commands