Remote Control Commands
R&S
®
NGL200/NGM200
110
User Manual 1178.8736.02 ─ 09
*RST:
2.050E+01
Parameters for setting and query:
<Channel list>
<list>
Example:
VOLT:ALIM:UPP? (@1)
Queries the upper safety limit for voltage at channel 1.
Example:
See
Example "Configuring the output voltage"
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit:LOWer
<current>
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit:LOWer?
[<Channel list>]
Sets or queries the lower safety limit for current.
Setting parameters:
<current>
<numeric value> | MIN | MINimum | MAX | MAXimum | <list>
<numeric value>
Numeric value for lower safety limit.
MIN | MINimum
Min value for lower safety limit.
MAX | MAXimum
Max value for lower safety limit.
Range:
For up to 6V: 0.001E+00 to 3.010E+00. For above
6V: 0.001E+00 to 6.010E+00
Increment: 0.001
*RST:
0.001E+00
Parameters for setting and query:
<Channel list>
<list>
Example:
CURR:ALIM:LOW? (@1)
Queries the lower safety limit for current at channel 1.
Example:
See
Example "Configuring the current output"
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]
<current>
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit[:UPPer]?
[<Channel list>]
Sets or queries the upper safety limit for current.
Setting parameters:
<current>
<numeric value> | MIN | MINimum | MAX | MAXimum | <list>
<numeric value>
Numeric value for upper safety limit.
MIN | MINimum
Min value for upper safety limit.
MAX | MAXimum
Max value for upper safety limit.
Range:
For up to 6V: 0.001E+00 to 3.010E+00. For above
6V: 0.001E+00 to 6.010E+00
Configuration Commands